Comparative
study of glucometer and laboratory glucose oxidase method for the estimation of
blood glucose levels in neonates
Naaz F.1,
Gulati R.K.2
1Dr. Farah Naaz,
Paediatrics, 2Dr. R.K.Gulati, Ex-Senior
Professor and HOD, both authors are attached with Department of Paediatrics, Government
Medical College Kota, Rajasthan, India.
Address For
Correspondence: Dr. Farah Naaz, Paediatrics,
Department of Paediatrics, Government Medical College Kota. Address- B 201
Golden Height Apartment, Flat No 502 Rajendra Marg, Bapunagar, Jaipur
(Rajasthan) India. E-mail: nfarah30@gmail.com
Abstract
Introduction:
Hypoglycemia is one of the most common
metabolic problems encountered in neonates. The glucose oxidase method used in
the laboratory for determining the blood glucose concentration is precise and
specific for glucose. The glucometers are often used for blood glucose
estimation in NICU. Objectives: This is a prospective study done to
determine the efficacy of glucometer in estimation of blood glucose levels in
neonates in comparison with the laboratory values. Methods: 170 neonates
admitted in NICU in Jay Kay Lon Hospital, Kota associated with Government Medical
College Kota with varied symptomatology were enrolled in this study. Blood
glucose estimation was done by glucometer and laboratory method using the same
venous sample at the time of admission. Statistical analysis was done by using
Pearson correlation. Hypoglycemia was defined as blood glucose level <45mg%.
Laboratory value was taken as gold standard. Results: Hypoglycemia was
common in neonates with risk factors like prematurity, meconium aspiration,
septicemia, birth asphyxia etc. In our study Glucometer had a specificity of
98.20%, sensitivity of 64.48 %, positive predictive value of 95%, negative
predictive value of 83.84% and accuracy of 86.47%. There was good correlation
and no significant difference between two methods in the range between 45 mg/dl
and 145 mg/dl but not at very high or very low glucose concentrations Conclusion:
The ‘Point of care “device - glucometer as a sole measuring device to
screen neonatal hypoglycemia is not satisfactory and confirmation with the
laboratory measurements of plasma glucose is still of up most importance.
Keywords:
Hypoglycemia, Neonates, Glucometer, Glucose
oxidase method
Author Corrected: 8th March 2019 Accepted for Publication: 14th March 2019
Introduction
Alteration
in blood glucose levels in newborns is difficult to detect clinically. Hence a
reliable “point of care”device (glucometer) for early detection and treatment
is needed. Hypoglycemia is one of the most common metabolic problems
encountered in the newborns. The overall incidence of hypoglycemia in neonates
varies from 0.2 to 11.4% [1,2]. However in the presence of certain risk factors
i.e. small for date, large for date, infants of diabetic mothers, prematurity
etc., the probability of hypoglycemia increases many folds[1]. Hypoglycemia in
neonates can be symptomatic and asymptomatic [3,4,5]. The most common symptoms
such as jitteriness, convulsions, apathy, hypotonia, coma, refusal to feed,
cyanosis, high pitched cry, hypothermia are very nonspecific and especially in
small sick infants, these symptoms may be easily missed. Therefore,
hypoglycemia must always be confirmed biochemically and by response to
treatment. Hypoglycemia is known to be associated with brain dysfunction and
neuromotor developmental retardation in both symptomatic and asymptomatic cases
[1,6].
In
symptomatic infant, plasma glucose concentration should be measured and if the
value is <45mg%, clinical interventions aimed at increasing blood glucose
concentrations are indicated [7].
The
methods for determining blood/plasma glucose concentration include
reductiometric method, glucose oxidase method and hexokinase method [1,8].The
glucose oxidase method used in the laboratory for determining the blood glucose
concentration is precise and specific for glucose [1]. As it is usually
performed in the main laboratory, the results are not available quickly enough
for timely appropriate management. The glucometers are often used for blood
glucose estimation in NICU. Many studies have shown that their results
co-relate well with the laboratory measured glucose levels in the normoglycemic
and hyperglycemic range but are not satisfactory in the lower range [9-16]. We
did a comparative study of blood glucose level in neonate using glucometer and
laboratory glucose oxidase method to determine the efficacy of glucometer in
comparison with the laboratory values.
Materials and Methods
Source of data: The
neonates admitted to NICU over a period of 3 months in Jay Kay Lon Hospital,
Kota associated with Government Medical College Kota.
Study design:
Prospective study
Sample size:170
Inclusion
criteria
1.Neonates
admitted in Jay Kay Lon Hospital Neonatal Intensive Care Unit (NICU)associated
with Government medical college, Kota.
2.Both
sex
Exclusion
criteria
1.Infant
>28 Days
2.Neonates
with PCV less than 40% more than 65%
3.Out
born or referral case
Method of collection of
data: Collection of data was by relevant
investigations i.e. glucose estimation by glucometer and laboratory glucose
oxidase method.
Methodology: At
admission, a detailed history was taken and a thorough physical examination was
performed so as to fulfill the inclusion and exclusion criteria laid down in
the study protocol.
Capillary
blood glucose levels measured using glucometer (ONE TOUCH-select simple) TM
with plasma obtained from heel prick using a stylet under aseptic
precautions. Simultaneously 2 ml of blood drawn from peripheral vein was
collected in fluoride tubes and sent to central clinical biochemistry
laboratory for plasma glucose measurement using glucose oxidase method.The
analysis was done on emergency basis within half an hour of collection of blood
sample. Routine quality control check was followed for the analysis. A
calibrated check using the manufacturer supplied control cuvette was performed
once a day on the glucose analyser.
Hypoglycemia
was defined as blood glucose levels less than 45 mg/dl Hyperglycemia was
defined as blood glucose levels more than 145 mg/dl .Blood glucose levels
45-145 mg/dl was considered as normal.Laboratory
value of glucose was taken as gold standard.
Statistical
analysis: Statistical methods used were
Descriptive Statistics, Frequencies, Contingency Coefficient Test (Crosstabs)
and correlation. All the statistical analysiswere carried out through the software
SPSS for Windows (version 16.0).
Results
170
neonates admitted in NICU of JKLON hospital, Kota were enrolled in our
study. Out of them 110 were males and 60 were females. 102 were term, 68 were
preterm and majority being less than 7 days. They had varied etiology and
clinical diagnosis of birth asphyxia, meconium aspiration syndrome/meconium
stained amniotic fluid, neonatal sepsis, neonatal hyperbilirubinemia,
respiratory distress syndrome and others (including preterm for nutritional and
supportive care, seizures for evaluation, premature rupture of membranes, acute
kidney injury, tracheo-esophageal fistula and other congenital malformations).
Neonates were classified according to their gestational age by WHO Fenton
charts. Maximum neonates (40.6%) were appropriate for gestational age (AGA). Remaining
17% were small for gestational age (SGA) and 2.4% were large for gestational age
(LGA).
Table
1,2 and figure 1 shows that there is a large correlation between the
glucometer and lab values when the blood glucose values, as determined by the
gold standard is >45mg%. But when the lab values (gold standard) are
<45mg% then there is just moderate correlation between the two. The accuracy of
glucometer to detect hypoglycemia in newborns in comparison with gold standard
(lab glucose oxidase method) is 86.47%.
Table-1: Diagnostic value of Glucometer in relation
to Lab value in detecting the hypoglycemia
Value |
Percentage |
Sensitivity |
64.40% |
Specificity |
98.20% |
PPV |
95% |
NPV |
83.84% |
Accuracy |
86.47% |
Table-2: Pearson
correlation: lab glucose oxidase method and
glucometer (capillary sample)
Lab Glucose Oxidase
Method (MG/DL) |
Total |
||||
<45 |
45-145 |
>145 |
|||
Glucometer (Capillary
Sample) In MG/DL |
<45 |
38 |
2 |
0 |
40 |
45-145 |
21 |
100 |
4 |
125 |
|
>145 |
0 |
1 |
4 |
5 |
|
Total |
59 |
103 |
8 |
170 |
RBC
Lab RBS
Figure-1:
Pearson correlation: lab glucose oxidase and glucometer (Capillary)
RBC
__ Linear
{RBC}
Contingency
coefficient(r) |
p
value |
+0.837 |
0.000 |
Discussion
Neonatal
hypoglycemia is a common metabolic disorder and the operational threshold
values of blood glucose less than 45 mg/dl should be used to guide
management.Thenormal range of blood glucose is variable and depends upon
factors like birth-weight, gestational age, body stores, feeding status,
availability of energy sources as well as the presence or absence of disease
[19,21]. All “at risk” neonates and sick infants should be monitored for blood
glucose levels.In our study we evaluated the efficacy of glucometers in
estimating the blood glucose levels in newborns, in comparison with laboratory
glucose oxidase method. Main stress was laid on detection of hypoglycemia by
both the methods and to know if glucometer is a good screening tool to detect
hypoglycemia. Laboratory glucose oxidase method of blood glucose estimation was
taken as gold standard. Hypoglycemia was
defined as blood glucose level less than 45mg% as defined by the study by Corn
lath et al[19].Our study showed that One Touch (select simple) glucometer had
sensitivity of 64.48% to detect hypoglycemia and specificity of 98.20 % to
detect hypoglycemia which is in correlation with other studies. This is in
accordance with Ngerncham S et al[17]
and Hamid MH et al[16] respectively(table
3).It was observed from our study that glucometer had a positive predictive
value of 95.00% which in contrast to Ngerncham S et al[17] and Ho HT et al[14]is
low and negative predictive value of 83.84 % which in contrast to
Ngerncham S et al[17] and Ho HT et al.is low[14](table 3).The accuracy of glucometer
to detect hypoglycemia in newborns in comparison with gold standard (lab
glucose oxidase method) is 86.47%.
Table-3: Comparison of sensitivity, specificity, PPV
& NPV of various studies
Sensitivity |
Specificity |
Positive
predictive value |
Negative
predictive value |
|
Our study |
64.40% |
98.20% |
95.00% |
83.84% |
Ngerncham S et al [17] |
62% |
100% |
100% |
85.7% |
Ho HT et al [14] |
46.7% |
79.2% |
73.7% |
54.3% |
Lab glucose oxidase
method is considered as gold standard. Although Glucometers are reported to
have several advantages in the analysis of glucose in newborns such as short
analysis time, small sample size, quick reading, and can be handled even by
unskilled personnel, still it remains as a screening tool. It has been shown
that most glucometers are inaccurate at very high or very low glucose
concentrations and certain variables like haematocrit, altitude, environmental
temperature or humidity and hypoxia may affect the result with bedside testing[15,20]
Correlation of glucometer and lab value: In
our study there was good correlation between two methods in the range between
45 mg/dl and 145 mg/dl. It has been shown that most glucometers are inaccurate
at very high or very low glucose concentrations and certain variables like
haematocrit, altitude, environmental temperature or humidity and hypoxia may
affect the result with bedside testing[13,15,17].
In
our study glucometer readings were higher in babies having glucose values below
45 mg/dl in relation to laboratory glucose oxidase method. Glucose values
measured using laboratory glucose oxidase method was higher in babies who had
glucose levels higher than 145 mg/dl when compared with glucometer. We observed
that few babies were overestimated for hypoglycemia when tested with
glucometer. So keeping in view the high specificity of glucometer in our study,
we conclude that, if the clinician feels neonate has hypoglycemia, but the
glucometer gives normal values, it is better to get the blood sugar repeated by
glucose oxidase method before giving any treatment.
Hence
our results agree with previous studies which concluded that glucose reagent
strips should be considered only as aeasy, rapid, point of care test and not as
a diagnostic test, due to their questionable reliability [13,15,18,22]
The
limitation of our study was firstly the small sample size in the higher
glucose range (>145 mg/dl). Further studies on a larger sample in higher
glucose range are needed. Secondly glucose estimation in our study did not
specify age in hours at the time of blood glucose estimation. Lastly whole
blood glucose level at room temperature in-vitro decline at a rate of 5-7
mg/dl/hour. The use of inhibitors such as fluoride, in collection tubes are
used to avoid this problem. The efficacy is best when prepared freshly. But the
fluoride tubes used in our study were pre-prepared commercially available
tubes.
Conclusion
There
is a variable detection rate of hypoglycemia by glucometer in the studies
mentioned and also in our study. Glucometer had good correlation at glucose
levels 45-145 mg/dl, still its
values in this study do not have sufficient validity to replace laboratory
testing in diagnosing hypoglycemia and hyperglycemia. We conclude that
glucometer test was satisfactory as aeasy, rapid, point of care test in diagnosing
neonatal hypoglycaemia but it can either over or underestimate glucose results.
Although the glucometer had good specificity and negative predictive value,
confirmation with laboratory measurements of plasma glucose and clinical
assessment of the infant are still of the utmost importance.
What this study adds to
our existing knowledge: Glucometers (point of
care devices) as a sole measuring device to screen neonatal hypoglycemia is not
satisfactory and confirmation with the laboratory measurements of plasma
glucose is still of up most importance.
Conflict of Interest:None
References
How to cite this article?
Naaz F, Gulati R.K. Comparative study of glucometer and laboratory glucose oxidase method for the estimation of blood glucose levels in neonates. Int J Pediatr Res. 2019;6(03):113-117.doi:10.17511/ijpr.2019.i03.02