Comparison of forehead infrared thermometry with axillary
digital thermometry in detecting neonatal hypothermia & study of effects of
early breast feeding and adequate clothing in maintaining neonatal temperature
Kurrey
V.1, Sao M.2, Phuljhele S.3
1Dr.Virendra
Kurrey, Associate Professor, 2Dr. Madhavi Sao, Postgraduate
fellow, 3Dr. Sharja Phuljhele, Professor & Head, all authors are
affiliated with the Department of Paediatrics, Pt. JNM
Medical College, Raipur, C.G., India.
Corresponding Author: Dr.
Madhavi Sao, A-84, Anandam World City, Kachna, Raipur Chhattisgarh, Email: saomadhavi@gmail.com
Abstract
Background:
Neonatal
hypothermia is a well-recognized important contributing factor to neonatal
morbidity and mortality especially in developing countries. Axillary
thermometry is the conventionalmethod in neonates to measure body temperature.
Forehead-infrared thermometry is new non-touch method that may reduce infection
rate and discomfort of neonates. Objectives:
To compare the accuracy of forehead infrared thermometry with axillary
digital thermometry in detecting neonatal hypothermia and to determine the
association of inadequate clothes and delayed initiation of breastfeeding with
increased risk of hypothermia. Materials
and Methods: A cross-sectional study was conducted among term neonates with
age < 24 hrs of life in the postnatal wards of tertiary care hospital. The
body temperature of neonates was measured by both methods & their accuracy
was analyzed to detect hypothermia with the adequate statistical method. Association
between delayed breastfeeding and inadequate clothing with neonatal hypothermia
were also evaluated. Result: Mean
difference (bias) of the axillary and infrared forehead readings (-0.29°C) in
the morning and (-0.31°C) in the evening. Agreement by the Bland-Altman method
in the morning (-0.76 & 1.33) and at evening (-1.33 & 0.76) shows the
transference of two techniques was inappropriate indicating that infrared-thermometry
cannot replace axillary digital thermometry. Subjects who were early breastfed and
with adequate clothing were found to have significantly higher body
temperature. Conclusion: Forehead
infrared thermometry cannot replace axillary thermometry and is not recommended
for neonatal temperature measurement. Early breastfeeding, covering of heads,
as well as extremities, were found to be protective from hypothermia.
Keywords: Neonatal, Hypothermia, Thermometry, Axillary,
Infrared
Author Corrected: 30th May 2019 Accepted for Publication: 7 th June 2019
Introduction
Body temperature
is an important vital sign in neonates. Prevention of neonatal hypothermia is
the most essential care in newborns. Accurate body temperature measurement is crucial
for detecting not only hyperthermia but also hypothermia for which neonates are
more vulnerable, so accurate temperature recording remains an essential
component of the neonatal care at birth and in the first few days of life [1]. Measurement of
temperature in neonates can be obtained by rectal, axillary, and tympanic
thermometry. Axillary temperature
measurement is recommended by the American Academy of Pediatrics and the
National Association of Neonatal Nurses. Mercury glass thermometer has been
replaced by a digital thermometer that is safer and more convenient.Furthermore
“minimal contact” being the guiding principle in neonatal care, more convenient
techniques have been evolved to replace axillary thermometry[2].
Among the new
methods, the non-contact forehead infrared thermometry is trendy in hospitals.
It is simple, fast and convenient compared to the conventional methods. Due to
its non-touch technique and negligible risk of cross infections, it appears to
be a promising method of thermometry in neonates where minimal handling is recommended[3]. In this
method, a sensor probe measures the amount of thermal radiation (infrared)
emitted from the forehead which has rich blood flow from the temporal artery[4].
Lack of
knowledge amongst health workers and mothers of simple methods to maintain the
warm chain from birth has been found to be the most common factor contributing
to hypothermia. Suitable policies comprising simple practices such as
establishing a warm delivery room, immediate drying at birth, skin to skin
contact, early breastfeeding, delay in bathing the newborn, proper clothing,
warm resuscitation, warm transportation, and training/ awareness, are necessary
to preventhypothermia[5].
Early and
adequate breastfeeding provides enough calories to newborns to generate body
heat. As up to 25% of heat loss can occur from an uncovered headso in the
initial hours after birth baby head and extremities should be adequately
covered[6].
In
this perspective, the present study aimedto compare the forehead infrared
thermometry with axillary digital thermometry in measuring neonatal temperature
for detection of hypothermia.The effects of early breastfeeding and covering
the heads and extremities on neonatal temperature were also studied and assessed
statistically for significance.
Materials and methods
Study
setting: This study was performed in the Paediatric
Department Pt. Jawaharlal Nehru Memorial Medical College & associated Dr.
B.R.AM Hospital Raipur, after approval from the institutional ethical
committee.
Type
of study: Prospective observational study conducted from
October 2017 to March 2018.
Sampling
method: A consecutive number of 400 apparently healthy term
neonates with age < 24 hrs of life were recruited for assessment of body temperature.
Inclusion
criteria: Term neonates less than 24 hours of age who were
kept with their mothers in postnatal wards.
Exclusion
criteria: Premature labor, prolonged rupture of membranes
[greater than 18 hours], gestational age < 37 weeks, perinatal asphyxia, signs
of illness or major congenital anomalies.
Study
methodology and data collection: After delivery, each
newborn was dried and placed under a radiant warmer. Mother and newborn shared
a single bed together in the postnatal ward. Intimate skin-to-skin contact
between mother and newborn was not practiced except during breast feeding. Newborns
body temperature was measured using an axillary digital thermometer (Omron
MC-670) and infrared thermometer (Omron MC - 720). Temperatures were recorded by both a digital
thermometer and infrared forehead thermometer once in the morning (10 am to 11 am)
and once in the evening (7 pm to 8 pm) after a shift to the postnatal ward. The
digital thermometer was put with the sensor in the newborn armpits after
activating the button power and waiting until the alarm sounds for its
temperature display. An Infrared
forehead thermometer was placed 1 to 3 cm from mid-forehead and the start
button was pressed for recording. The measurement was completed in 1 second
with a long beep.
The data was
collected using an especially designed proforma having details of postnatal age
of the newborns in hours, gestational age, birthweight, recorded temperatures
of the newborn, received early breastfeeding or not, head/extremities covered
or not and temperature of the postnatal ward.
Statistical
analysis
·
Data were expressed in percentage and
Mean.
·
To study the statistical agreement
between two methods, the Lin Concordance correlation coefficient and a newer
method of comparison Bland and Altman plot method were applied.
·
The Student's t-test was used to check
the significance of mean difference parametric data.
·
A mean difference of temperature ±0.5°C
was considered clinically acceptable.
Results
The
body temperature of 400 neonates was measured including 207 (51.75%) males and
193 (48.25%) females. The mean birth weight was 2840±359 grams. 359 (89.8%)
subjects had birth weight 2.5-4.2 Kg while, 41 (10.2%) subjects had birth
weight 1.5-2.499 kg. The temperature of all the neonates was measured by both
the techniques in the morning as well as evening.
Table-1:
Comparison of IR forehead and digital axillary temperature in neonates
Temperature |
Morning and Evening IR forehead temperature in neonates |
Morning and Evening digital axillary temperature in neonates |
||
Morning |
Evening |
Morning |
Evening |
|
Moderate hypothermia 32.0-35.9
°C |
7.50% |
3.80% |
14.20% |
14.20% |
Mild hypothermia 36.0-36.4°C |
4.20% |
4% |
31.50% |
40% |
Normal
36.5-37.5°C |
88.20% |
92.20% |
54.20% |
45.80% |
Total |
100 |
100 |
100 |
100 |
Morning
and evening IR forehead temperature in the study subjects was assessed; Table 1
shows that in the morning 11.70% neonates and in the evening 7.80% were in hypothermia.
Hypothermia was detected by digital axillary thermometer in45.70% of newborns
in the morning and 44.20% newborns in the evening. The difference in percentages
of morning and evening hypothermia cases shows that byIR forehead temperature
measurement 34% and 36.40% neonates were less detected for hypothermia in comparison
to a digital axillary method.
Table-2: Mean, range, mean
difference and limits of agreement of the temperaturemeasurements
Variable |
The Axillary temperature in °C |
The infrared temperature in °C |
Mean difference in °C |
The lower limit of agreement |
The upper limit of agreement |
p-value |
||
Mean |
Range |
Mean |
Range |
|||||
Morning |
36.48 |
35.5-38.7 |
36.77 |
34.6-38 |
-0.29 |
-0.7627 |
1.3388 |
<0.0001 |
Evening |
36.52 |
34.8-37.8 |
36.83 |
35.8-38.3 |
-0.31 |
-1.3388 |
0.7627 |
<0.0001 |
The mean, range, mean difference and 95%
limits of agreement of the temperature measurements are shown in Table 2.
Considering all the 400 pair of readings, the mean axillary temperature at
morning was 36.48°C and the mean infrared temperature at morning was 36.77°C. The
mean axillary temperature in the evening was 36.83°C and the mean infrared
temperature in the evening was 36.83°C. Therefore, the mean infrared
temperature recorded was significantly higher than the mean axillary
temperature (p-value <0.0001).
The mean
difference (bias) of the axillary and infrared forehead readings in the morning
was -0.29°C and at evening was -0.31°C. This is less than the clinically
acceptable value set at ± 0.5°C. Infrared readings tend to be a little greater
than the axillary readings. The lower and upper limits of the agreement by the
Bland-Altman method in the morning were -0.7627&1.3388, and at evening
-1.3388 & 0.7627 respectively (Figure 1& 2). This range of the limits
of around 2°C is too wide to be clinically acceptable.
Figure-1:Bland
Altman plot for comparability of IR forehead temperature and
digital
axillary temperature in the morning
Figure
2: Bland Altman plot for comparability of IR forehead temperature and
digital
axillary temperature in the evening
Correlation
analysis between morning and evening IR forehead temperature and the digital
axillary temperature was performed using Pearson correlation analysis. Table 3
shows that only mild correlation exists between two tests making transference
of the technique of measurement of temperature inappropriate.
Table-3: Correlation between IR
forehead temperature and digital axillary temperature
Variable |
Correlation coefficient r |
P value |
Interpretation |
|
Morning IR forehead temperature |
Morning digital axillary
temperature |
0.567 |
<0.0001 |
Mild upstream correlation |
Evening IR forehead temperature |
Evening digital axillary
temperature |
0.374 |
<0.0001 |
Mild upstream correlation |
Lin’s
Concordance correlation coefficient between IR forehead temperature and digital
axillary temperature in morning and evening was calculated. The sample
concordance correlation coefficient (ρc) for the morning was 0.4834 (0.42-0.53)
and for the evening (ρc) was 0.4933 (0.44-0.54) indicating a moderate degree of
concordance, thus rendering transference of technique is inappropriate.
Table-4: Effect of head covering on
morning and evening temperature
|
Head covered morning |
N |
Mean |
S.D. |
S.E. of mean |
T |
p-value |
Morning digital axillary temp
of the newborn (°C) |
Yes |
210 |
36.6724 |
0.50727 |
0.035 |
8.26 |
<0.0001 |
No |
190 |
36.28 |
0.43476 |
0.03154 |
|||
Evening digital axillary temp
of the newborn (°C) |
Yes |
219 |
36.6329 |
0.51126 |
0.03455 |
4.63 |
<0.0001 |
No |
181 |
36.3978 |
0.4971 |
0.03695 |
Table
4 shows the effect of head covering on the morning (mean difference 0.39°C)
and evening temperature (mean difference 0.24°C)
was performed using Student's unpaired t-test. A significant difference
(p<0.001) was observed between two temperatures indicating that temperature
was significantly higher when the head was covered.
Table-5: Effect of extremities
covering on morning and evening temperature
|
Extremities covered morning |
N |
Mean |
S.D. |
S.E. of mean |
t |
p-value |
|
Morning digital axillary temp
of the newborn (°C) |
Yes |
200 |
36.709 |
0.51346 |
0.03631 |
9.65 |
<0.0001 |
|
No |
200 |
36.263 |
0.40416 |
0.02858 |
||||
Evening digital axillary temp
of the newborn (°C) |
Yes |
285 |
36.6558 |
0.52341 |
0.031 |
8.54 |
<0.0001 |
|
No |
115 |
36.2061 |
0.332 |
0.03096 |
Table
5 shows the effect of extremities covering on the morning (mean difference 0.45°C)
and evening temperature (Mean difference 0.45°C)
was performed using Student's unpaired t-test.
A significant difference was observed between two temperatures
indicating that temperature was significantly higher when the head was covered.
Table-6: Effect of early breastfeeding on temperature
Variable
|
Early
Breastfeeding within 1 hr |
N |
Mean |
S.D. |
S.E.M. |
t |
P-value |
Morning
digital axillary temp of the newborn (°C) |
Yes |
186 |
36.5183 |
0.59091 |
0.04333 |
1.175 |
0.241 |
No |
214 |
36.4579 |
0.43282 |
0.02959 |
|
|
|
Evening
digital axillary temp of the newborn (°C) |
Yes |
186 |
36.6667 |
0.54621 |
0.04005 |
5.211 |
<0.0001 |
No |
214 |
36.4047 |
0.45921 |
0.03139 |
|
|
In
Table 7 effect of early breastfeeding on temperature was assessed using
Student's unpaired t-test. While early breastfeeding did not impact morning
temperature, subjects who were early breastfed were found to have significantly
higher body temperature in the evening.
Discussion
Infrared
thermometry is a new method compared to conventional methods of temperature
measurement which is rapid and easy to use.
As it is a non-touch technique there is a negligible chance of cross
infections in neonates and very suitable to use. Various studies have been conducted to
compare its accuracy with the axillary method of temperature measurement.
In the present
study, a high value of bias and wide limits of agreement were noted. The mean
difference (bias) of the axillary and infrared forehead readings in the morning
was -0.29 °C and at evening was -0.31°C. This is less than the clinically
acceptable value set at ±0.5°C. Infrared readings tend to be a little greater
than the axillary readings. The lower and upper limits of the agreement by the
Bland-Altman method in the morning were -0.7627 & 1.3388, and at evening
-1.3388 & 0.7627 respectively (Table 2). This range of the limits of around
2°C is too wide to be clinically acceptable.
Among
the studies comparing axillary and forehead infrared thermometry,Sethi et al. 2013noted a mean difference of
-0.5 C and limits of agreement as -2.3, 1.2 in their study comparing axillary
and forehead infrared thermometry[7].Megha S Patel
et al. 2014 found a very high mean difference of -1.5 C and 95% limits of
agreement as -2.7,-0.3 with infrared values higher than the axillary values in
all the pairs of readings[8]. Ilaria Merusiet al 2015 recorded a mean
difference of 0.35 C with 95% limits of agreement as -0.45, 1.17[9] and Uslu et al. 2011 noted a bias of -0.55 C[10]. In contrast, Chiappini et al. 2011 reported a good agreement
(mean difference = 0.070C, 95% limits of agreement: [-0.62, 0.76]) between
Infrared forehead thermometry (IRFT) and axillary thermometry using glass
mercury thermometer in paediatric population[11].Kotsia et al. 2015reported that poor
correlation existed between forehead temperature and digital axillary
temperature[12].
They
all concluded that infrared non-contact thermometers cannot be recommended for
the measurement of body temperature in neonates in an intensive care setting
where accurate temperature measurement is required.
In
the present study, the effect of head covering on the morning (mean difference
0.39°C) and evening temperature (mean difference 0.24°C) was assessed and the
significant difference was observed (p<0.001 between two temperatures
indicating that temperature was significantly higher when the head was covered.
Also, the effect of extremities covering on the morning (mean difference
0.45°C) and evening temperature (mean difference 0.45°C) shows a significant
difference was (p<0.001) between two temperatures indicating that temperature
was significantly higher when the head was covered.
McCall et al. 2006 stressed the
importance of covering the head and extremities of babies in order to prevent
hypothermia[13].Delavar et al. 2014 noted that hypothermia was
not found to be associated with keeping mother and baby together, skin-to-skin
contact, gestational age, delayed appropriate clothing, breastfeeding[14]. In the present
study who were early breastfed were found to have significantly higher body
temperature (OR =2.78).Studies suggesting similar findings have observed that
breastfeeding is found to be protective against hypothermia and delayed
breastfeeding >24 hours after birth is associated with increased risk of
neonatal hypothermia (OR 3.32) [15, 16].
Conclusion
Infrared
forehead thermometer readings do not comparably correspond to axillary digital
thermometer& infrared forehead thermometer may miss the diagnosis of
hypothermia in the neonate.It was also observed that the covering of head and
extremities and early breastfeeding are protective against hypothermia. Hereby
it is recommended infrared forehead thermometer may not be useful for
temperature monitoring in neonates.
References
How to cite this article?
Kurrey V, Sao M, Phuljhele S. Comparison of forehead infrared thermometry with axillary digital thermometry in detecting neonatal hypothermia & study of effects of early breast feeding and adequate clothing in maintaining neonatal temperature. Int J Pediatr Res. 2019; 6(06):286-291.doi:10.17511/ijpr.2019.i06.04