Light-emitting Diodes versus Compact Fluorescent Tubes for Phototherapy in Neonatal Jaundice: A randomised control trial

  • Dr. T. Raghunath Reddy Aditya Hospital Hyderabad, Telangana
  • Dr. P. Krishna Prasad Aditya Hospital Hyderabad, Telangana
  • Dr. Hemant parakh Aditya Hospital Hyderabad, Telangana
  • Dr. Pritesh nagar Aditya Hospital Hyderabad, Telangana
Keywords: Phototherapy, Neonatal jaundice, Neonatal hyyperbilirubinemia

Abstract

Introduction: Newer light emitting diodes (LED) phototherapy units have become indigenous part in the management of neonatal hyperbilirubinemia almost replacing the conventional compact fluorescent light (CFL) units. Only few studies from India available compare the efficacies of two, so a randomised control study was undertaken.

Methods: 60 healthy neonates ≥ 35 weeks gestational age were randomised into CFL and LED groups using computerised random number sequence. The rate of fall of bilirubin levels at 6 hours and at completion along with total duration of phototherapy in both groups was measured. Results analysed by standard statistical methods.

Results: LED phototherapy units showed higher rate of fall bilirubin at 6hrs (LED 0.41± 0.10mg/dl/hr VS CFL 0.23± 0.17 mg/dl/hr, C.I=95%, p<0.05) and after completion of therapy (LED 0.29± 0.08 mg/dl/hr VS CFL 0.19± 0.06 mg/dl/hr with p<0.05 C.I 95%) compared to CFL group. There was no statistically significant difference in total duration of therapy in both the groups (LED 18.8±8hrs VS CFL 20±6 hrs C.I=95% p=0.56). 2 among CFL group had failure of phototherapy whereas 1 was lost to follow up in LED group. Both groups didn’t show any side effects.

Conclusion: The LED phototherapy units are more efficacious in terms of higher rate of fall of bilirubin levels in similar time duration compared to CFL units. But the total duration of phototherapy was equal; it can be because there was significant difference between both the groups in terms of bilirubin levels at start and at end of therapy which couldn’t be prevented even after effective randomisation.

Downloads

Download data is not yet available.

References

Kliegman RM, Behrman RE, Jenson HB, Stanton BF. Nelson Textbook of Paediatrics 18th edition. Saunders Elsevier Publications 2007;102.3: 756-772

Mccandles, David W. History of Jaundice –Kernicterus, Contemporary Clinical Neurosciences 14th edition. Humana press2011: 286-290.

Baty JM, Blackfan KD & Diamond LK. Blood studies in infants and in children. I. Erythroblastic anemia; a clinical and pathologic study. American Journal of Diseases in Childhood 1932;43(3):667-704.

Hansen TW. Pioneers in the scientific study of neonatal jaundice and kernicterus. Pediatrics 2000; 106(2):E15.

Brown AK, Kim MH, Wu PY, Bryla DA. Efficacy of phototherapy in prevention and management of neonatal hyperbilirubinemia. Pediatrics. 1985 Feb;75(2 Pt 2):393-400.

Tan KL, Lim GC, Boey KW. Efficacy of “highintensity” blue-light and “standard” daylight phototherapy for non haemolytic hyperbilirubinaemia. Acta Paediatr. 1992 Nov;81(11):870-4.

Cremer RJ, Perryman PW, Richards DH. Influence of light on the hyperbilirubinaemia of infants Lancet. 1958 May 24;1(7030):1094-7.

Ostrow JD. Mechanism of bilirubin photo degradation. Semin Hematol. 1972 Apr;9(2):113-25.

Lightner et al. Bilirubin photooxidation products in the urine of jaundiced neonates receiving phototherapy; Pediatric Research 1984;18(8):696-700.

Prates et al. Laser investigation of bilirubin photobilirubin photoconversion. Pediatr Res. 1985 Feb;19(2):166-71.

Ennever JF. Blue light, green light, white light, more light: treatment of neonatal jaundice. Clin Perinatol. 1990 Jun;17(2):467-81.

Agarwal R, Deorari A, Paul VK. Jaundice in newborn- AIIMS Protocols in Neonatology 1st edition CBS publishers 2015; 131-133

Subramanian, et al. Evaluation Of phototherapy devices. Indian Pediatr. 2011 Sep;48(9):689-96. Epub 2010 Nov 30.

Maisels MJ, Kring EA, DeRidder J. Randomized controlled trial of light-emitting diode phototherapy.Journal of Perinatology 2007;27(9):565–567.

Seidman DS, Moise J, Ergaz Z, Laor A, Vreman HJ, Stevenson DK, et al. A new blue light-emitting phototherapy device: A prospective randomized controlled study. Journal of Pediatrics 2000;136(6):771–774.

Tan KL. Efficacy of bidirectional fiberoptic phototherapy for neonatal hyperbilirubinemia. Pediatrics 1997;99(5):E13.

Seidman DS, Moise J, Ergaz Z, et al. A prospective randomized controlled study of phototherapy using blue and blue-green light-emitting devices, and conventional halogen-quartz phototherapy .J Perinatol. 2003 Mar;23(2):123-7.

Rosen D, Rosen A. inventors; Therapeutic Method and Internally Illuminated Garment for the Management Disorders Treatable by Phototherapy. US Patent Number 6045575. 2000. Available at: www.google.com/patents/US6045575.

Rosen H, Rosen A, Rosen D, Onaral B, Hiatt M. Use of a Light Emitting Diode (LED) array for bilirubin phototransformation. Conf Proc IEEE Eng Med Biol Soc. 2005;7:7266-8.

Vreman HJ, Wong RJ, Stevenson DK, Route RK, Reader SD, Fejer MM, Gale R & Seidman DS. LightEmitting Diodes: A novel light source for phototherapy. Pediatric Research1998; 44(5): 804–809.

Agarwal R, Kaushal M, Aggarwal R et al. Early neonatal hyperbilirubinemia using first day serum bilirubin level. Indian Pediatrics2002; 39(8):724-730.

Maisels MJ, Baltz RD, Bhutani V, Newman TB, Palmer H, Rosenfeld W et al. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics2004; 114: 297–316.

American Academy of Pediatrics Subcommittee on Hyperbilirubinemia. Management of hyperbilirubinemia in the newborn infant 35 or more weeks of gestation. Pediatrics. 2004 Jul;114(1):297-316.

Bhutani V, Gourley GR, Adler S, Kreamer B, Dalman C, Johnson LH. Noninvasive measurement of total serum bilirubin in a multiracial predischarge newborn population to assess the risk of severe hyperbilirubinemia. Pedia- trics 2000; 106(2):E16.

Kumar P, Jain N Thakre R, Murki S, Venkataseshan S. Management of hyperbilirubinemia. Evidence based clinical practice guidelines-. National Neonatal Forum of India. October 2010: 139-151.

Olav, Kallenburg. Foundations of Modern Probability. Spinger series in statistics 2nd edition 2002;650 .

Kumar P, Murki S, Malik GK, Chawla D, Deorari AK, Karthi N, et al. Light emitting diodes versus compact fluorescent tubes for phototherapy in neonatal jaundice: a multi center randomized controlled trial. Indian Pediatrics 2010;47(2):131–137.

Kumar P, Chawla D, Deorari A. Light-emitting diode phototherapy for unconjugated hyperbilirubinaemia in neonates. Cochrane Database Syst Rev. 2011 Dec7;(12):CD007969. doi:10.1002/14651858.CD007969.pub2..

Martins BM, Carvalho MD, Moreira ME, Lopes JM. Efficacy of new microprocessed phototherapy system with five high intensity light emitting diodes (Super LED). Journal of Pediatrics (Rio J) 2007;83(3):253–258.

Karagöl BS, Erdeve O, Atasay B, Arsan S. Efficacy Of Light Emitting Diode phototherapy in comparison to conventional phototherapy in neonatal jaundice. Journal of Ankara University Faculty of Medicine 2007; 60(1):31-34.

Colindres V, Rountree C ,Oden M et al. Prospective randomized controlled study comparing low-cost LEDand conventional phototherapy for treatment of neonatal hyperbilirubinemia. Journal of Tropical Pediatrics 2011;58(03):178-183.

Light-emitting Diodes versus Compact Fluorescent Tubes for Phototherapy in Neonatal Jaundice: A randomised control trial
CITATION
DOI: 10.17511/ijpr.2014.i03.02
Published: 2014-12-31
How to Cite
Dr. T. Raghunath Reddy, Dr. P. Krishna Prasad, Dr. Hemant parakh, & Dr. Pritesh nagar. (2014). Light-emitting Diodes versus Compact Fluorescent Tubes for Phototherapy in Neonatal Jaundice: A randomised control trial. Pediatric Review: International Journal of Pediatric Research, 1(3), 72-79. https://doi.org/10.17511/ijpr.2014.i03.02
Section
Original Article