Comparative study of glucometer and laboratory glucose oxidase method for the estimation of blood glucose levels in neonates

  • Dr. Farah Naaz Department of Paediatrics, Government Medical College Kota, Rajasthan, India.
  • Dr. R.K.Gulati Department of Paediatrics, Government Medical College Kota, Rajasthan, India.
Keywords: Hypoglycemia, Neonates, NeonatesGlucose oxidase method

Abstract

Introduction: Hypoglycemia is one of the most common metabolic problems encountered in neonates. The glucose oxidase method used in the laboratory for determining the blood glucose concentration is precise and specific for glucose. The glucometers are often used for blood glucose estimation in NICU.

Objectives: This is a prospective study done to determine the efficacy of glucometer in estimation of blood glucose levels in neonates in comparison with the laboratory values.

Methods: 170 neonates admitted in NICU in Jay Kay Lon Hospital, Kota associated with Government Medical College Kota with varied symptomatology were enrolled in this study. Blood glucose estimation was done by glucometer and laboratory method using the same venous sample at the time of admission. Statistical analysis was done by using Pearson correlation. Hypoglycemia was defined as blood glucose level <45mg%. Laboratory value was taken as gold standard.

Results: Hypoglycemia was common in neonates with risk factors like prematurity, meconium aspiration, septicemia, birth asphyxia etc. In our study Glucometer had a specificity of 98.20%, sensitivity of 64.48 %, positive predictive value of 95%, negative predictive value of 83.84% and accuracy of 86.47%. There was good correlation and no significant difference between two methods in the range between 45 mg/dl and 145 mg/dl but not at very high or very low glucose concentrations

Conclusion: The ‘Point of care “device - glucometer as a sole measuring device to screen neonatal hypoglycemia is not satisfactory and confirmation with the laboratory measurements of plasma glucose is still of up most importance.

Downloads

Download data is not yet available.

References

1. Singhal PK, Singh M, Paul VK, et al. Neonatal hypoglycemia--clinical profile and glucose requirements. Indian Pediatr. 1992 Feb;29(2):167-71.[pubmed]

2. Williams AF. Hypoglycaemia of the newborn: a review. Bull World Health Organ. 1997;75(3):261-90.[pubmed]

3. Charles AS &Eugina KP: Disorders of Carbohydrate Metabolism. In: Avery Disease of the Newborn. Teuscch HW, Ballard RA & Gleason CA. 8TH ed. Philadelphia: SAUNDERS, 2005; 1410-1420.

4. Kleigman RM: Problems In Metabolic Adaptation: Glucose, Calcium and Magnesium. In: Care Of the High Risk Neonate. Klaus MH &Faranoff AA. 5th ed. Philadelphia: SAUNDERS, 2000; 304-309.

5. Barbara JS and Robert MK: The Endocrine System. In: Nelson Text Book of Pediatrics. Behrman RE, Kliegman RM, Jenson HB. 17TH ed. Philadelphia: SAUNDERS, 2004; 614-616.

6. Alkalay AL, Flores-Sarnat L, Sarnat HB, et al. Brain imaging findings in neonatal hypoglycemia: case report and review of 23 cases. Clin Pediatr (Phila). 2005 Nov-Dec;44(9):783-90. DOI:10.1177/000992280504400906.[pubmed]

7. Cornblath M, Hawdon JM, Williams AF, et al. Controversies regarding definition of neonatal hypoglycemia: suggested operational thresholds. Pediatrics. 2000 May;105(5):1141-5.[pubmed]

8. Gregory A. Threatte, John Bernard henry: Carbohydrates. In: Clinical Diagnosis and management by laboratory methods. JB Henry. 19th ed. Noida: Harcourt Asia PTE Ltd, 1999; 194-207

9. Narayan S, Aggarwal R, Deorari AK, et al. Hypoglycemia in the newborn. Indian J Pediatr. 2001 Oct;68(10):963-5.[pubmed]

10. Michel A, Küster H, Krebs A, et al. Evaluation of the Glucometer Elite XL device for screening for neonatal hypoglycaemia. Eur J Pediatr. 2005 Nov;164(11):660-4. Epub 2005 Jul 22. DOI:10.1007/s00431-005-1733-9.[pubmed]

11. Pamela Kirkham, Andrew Watkins. Comparison of two reflectance photometers in the assessment of neonatal hypoglycemia. Arch Dis Child Fetal Neonatal Ed 1995; 73: F170-73.

12. Dahlberg, Marianne, et al. Comparison of a new instant blood glucose method with conventional laboratory analysis in diagnosis of neonatal hypoglycemia. Pediatr Res 1996; 39(4)307.

13. Mehta, Nilesh M, Whitelaw, Andrew, Stone, Janet. Diagnosis of neonatal hypoglycemia at the cot side. A comparison of precision Q.I.D., Hemocue and Hexokinase Methods. Pediatr Res 1999; 45(94): 211A.

14. Ho HT, Yeung WK, Young BW. Evaluation of "point of care" devices in the measurement of low blood glucose in neonatal practice. Arch Dis Child Fetal Neonatal Ed. 2004 Jul;89(4):F356-9. DOI:10.1136/adc.2003.033548

15. Ellis M, Manadar DS, Manadar N, JM Land, NPatel, AM de L Costello. Comparision of two cotside methods for detection of hypoglycemia among neonates in Nepal. Arch Dis Child Fetal Neonatal Ed 1996; 75: F122-5.

16. Hamid MH, Chishti AL, Maqbool S. Clinical utility and accuracy of a blood glucose meter for the detection of neonatal hypoglycemia. J Coll Physicians Surg Pak. 2004 Apr;14(4):225-8. DOI:04.2004/JCPSP.225228

17. Ngerncham S, Piriyanimit S, Kolatat T, et al. Validity of two point of care glucometers in the diagnosis of neonatal hypoglycemia. Indian Pediatr. 2012 Aug;49(8):621-5. Epub 2011 Jan 17.[pubmed]

18. Arun kumar De, Biswas R, Samanta M. Study of blood glucose level in normal and low birth weight newborns and impact of early breast feeding in a tertiary care centre. Ann Nigerian Med 2011;5:53-8..

19. Cornblath M, Ichord R. Hypoglycemia in the neonate. SeminPerinatol. 2000 Apr;24(2):136-49.[pubmed]

20. Kilpatrick ES1, Rumley AG, Myint H, et al. The effect of variations in haematocrit, mean cell volume and red blood cell count on reagent strip tests for glucose. Ann Clin Biochem. 1993 Sep;30 ( Pt 5):485-7. DOI:10.1177/000456329303000513.[pubmed]

21. Mitanchez D. Glucose regulation in preterm newborn infants. Horm Res. 2007;68(6):265-71. Epub 2007 Jun 20. DOI:10.1159/000104174.[pubmed]

22. Harris DL, Battin MR, Weston PJ, et al. Continuous glucose monitoring in newborn babies at risk of hypoglycemia. J Pediatr. 2010 Aug;157(2):198-202.e1. doi: 10.1016/j.jpeds.2010.02.003. Epub 2010 Mar 24.[pubmed]
Comparative study of glucometer and laboratory glucose oxidase method for the estimation of blood glucose levels in neonates
CITATION
DOI: 10.17511/ijpr.2019.i03.02
Published: 2019-03-31
How to Cite
Dr. Farah Naaz, & Dr. R.K.Gulati. (2019). Comparative study of glucometer and laboratory glucose oxidase method for the estimation of blood glucose levels in neonates. Pediatric Review: International Journal of Pediatric Research, 6(3), 113-117. https://doi.org/10.17511/ijpr.2019.i03.02
Section
Original Article